Kett Fd 800 Dual Control Moisture Balance
Kett Fd 800 Dual Control Moisture Balance
FD800 Dual Control Moisture Balance
Flagship Moisture Balance Dual Temperature Control for Complex, Scorch-Prone Samples
The top of Kett’s loss-on-drying line, built for the materials that defeat ordinary balances: liquids, pastes, powders, and large particulates. A second, non-contact sensor reads the sample surface directly, so heat-sensitive and volatile samples dry fast without burning. On the proven FD720 platform, including the large 120 g capacity, six drying modes, patented auto-taring, the FD800 handles a demanding mix of materials with primary-method accuracy.
- Dual temperature control: A chamber thermistor plus a non-contact radiation thermometer that controls to the sample surface up to 250 °C. This is aggressive heat for speed, eased down as the surface dries.
- Predictive measuring mode: Compares projected against actual weight change and ends the test once results meet confidence limits resulting in maximum throughput without cutting a reading short.
- Guaranteed accuracy from a primary method: Loss on drying measures mass loss directly; run to completion, the result is definitive and stands up to scrutiny.
- Stored profiles and results memory: Ten stored measuring conditions for instant recall across materials, plus on-board memory for 100 results.
Specifications :
| Model | FD800 |
| Measurement Range | 0-100%WB 0-500% DB |
| Measurement Indication | .01% or 1mg |
| Accuracy | 0.02-10g sample 0.05-5g sample |
| Display | Moisture/Weight/Measure-ment Time/Temperature |
| Drying Time | To 240 Minutes. |
| Measurement Weight | 0.1-120g |
| Sample Pan | 130mm Diameter 13mm Deep |
| Temperature Range | 30-180C° (T1) 30-250C° (T2) |
| Power Source | AC100-120/220-240 |
| Measuring Principle | Weight Loss by Heat (LOD) |
| Heat Source | Mid-IR Quartz |
| Display | Backlit LCD (137x43mm) |
| Output | RS232C Printer/PC |
| Measurement Modes | 6 |
| Test Protocol Storage | 100 |
| Measurement Storage | 100 Tests |
| Power Consumption | 900W |
| Weight (Net/Ship) | 5.4/9kg |
| Dimensions | 220(w)x415(d)x220(h) |
| Warranty | 3 Years Parts and Labor Post Warranty Maintenance Agreements available - con-tact us for details. |
FD800 Dual Control Moisture Balance
Flagship Moisture Balance Dual Temperature Control for Complex, Scorch-Prone Samples
The top of Kett’s loss-on-drying line, built for the materials that defeat ordinary balances: liquids, pastes, powders, and large particulates. A second, non-contact sensor reads the sample surface directly, so heat-sensitive and volatile samples dry fast without burning. On the proven FD720 platform, including the large 120 g capacity, six drying modes, patented auto-taring, the FD800 handles a demanding mix of materials with primary-method accuracy.
- Dual temperature control: A chamber thermistor plus a non-contact radiation thermometer that controls to the sample surface up to 250 °C. This is aggressive heat for speed, eased down as the surface dries.
- Predictive measuring mode: Compares projected against actual weight change and ends the test once results meet confidence limits resulting in maximum throughput without cutting a reading short.
- Guaranteed accuracy from a primary method: Loss on drying measures mass loss directly; run to completion, the result is definitive and stands up to scrutiny.
- Stored profiles and results memory: Ten stored measuring conditions for instant recall across materials, plus on-board memory for 100 results.
Specifications :
| Model | FD800 |
| Measurement Range | 0-100%WB 0-500% DB |
| Measurement Indication | .01% or 1mg |
| Accuracy | 0.02-10g sample 0.05-5g sample |
| Display | Moisture/Weight/Measure-ment Time/Temperature |
| Drying Time | To 240 Minutes. |
| Measurement Weight | 0.1-120g |
| Sample Pan | 130mm Diameter 13mm Deep |
| Temperature Range | 30-180C° (T1) 30-250C° (T2) |
| Power Source | AC100-120/220-240 |
| Measuring Principle | Weight Loss by Heat (LOD) |
| Heat Source | Mid-IR Quartz |
| Display | Backlit LCD (137x43mm) |
| Output | RS232C Printer/PC |
| Measurement Modes | 6 |
| Test Protocol Storage | 100 |
| Measurement Storage | 100 Tests |
| Power Consumption | 900W |
| Weight (Net/Ship) | 5.4/9kg |
| Dimensions | 220(w)x415(d)x220(h) |
| Warranty | 3 Years Parts and Labor Post Warranty Maintenance Agreements available - con-tact us for details. |